Centrifuge Module Interface Operating Instructions Introduction
092D0001-02 Rev. A page 27
The Centrifuge module Settings screen includes the following areas,
fields, and buttons:
Table 2: LineMaster Settings Screen Fields
Area Description
Enter the rotation
speed or
Centrifuge force
This area includes the following fields:
Target RPMs (revolutions per minute) specifies
the rotational speed of the Hettich unit.
Target RCF (relative centrifugal force) specifies
the centrifugal force of the Hettich unit.
Entering a value in the Target RPMs or Target RCF
fields causes the other field to automatically update
to the corresponding value.
Enter a Spin Time This area includes the following fields:
Spin Time (sec) Controls the time in seconds
spent at the rotational speed. The default value
is 595 seconds (9 minutes, 55 seconds); you can
enter a value here to modify the spin time.
Acceleration (sec) The system calculates this
value from the Target RPM or Target RCF
fields. You cannot modify this field.
Deceleration (sec) The system calculates this
value from the Target RPM or Target RCF
fields. You cannot modify this field.
Cycle Time (sec) The sum of the Spin Time,
Acceleration, and Deceleration values. The
default is 722 seconds (12 minutes, 2 seconds).
You cannot modify this field.
Hettich Loading
Timers
This area includes the following fields:
Load Delay (sec) Specifies the time in seconds
that the robot waits between the arrival of a
sample tube in the Centrifuge module and
placing the cassette containing that sample
tube into the Hettich unit.
Tube Arrival (sec) Specifies the maximum
amount of time between consecutive sample
tubes arriving at the interface gate before
loading cassettes into the Hettich unit.