Troubleshooting Centrifuge Module Interface Operating Instructions
page 82 092D0001-02 Rev. A
b. If you did not remove the sample tube, select No.
Inspect the Centrifuge module for the missing tube. If you find the
tube, you can manually remove it from the Centrifuge module.
4. Repeat these steps for all missing sample tubes.
Cleaning after Sample Tube Breakage
After removing a broken tube, use the cleaning procedures in Chapter 3,
Maintenance to clean all affected areas of the Hettich unit and Centrifuge
module, as necessary:
Cleaning the Hettich Unit Outer Enclosure , page 59
Cleaning the Hettich Unit Chamber and Seals , page 63
Inspecting and Cleaning the Buckets and Rotor, page 60
Cleaning the Cassettes, page 51
Cleaning the Scale, page 50
Cleaning the Robot Gripper, page 47
Cleaning the Interface Gate Tube Holder, page 53
BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to
Appendix A, Safety Information for recommended precautions when
working with biohazardous materials.