Siemens ADVIA LabCell - Understanding E-stop Conditions; Lifting the Robot from the Hettich Unit

Siemens ADVIA LabCell
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Centrifuge Module Interface Operating Instructions
page 72 092D0001-02 Rev. A
2. Turn the Centrifuge module on.
See Turning the Centrifuge Module On‚ page 35.
E-stop Conditions
An error at the Centrifuge module can trigger an E-stop. When a station
E-stop condition occurs, the following actions take place:
•The robot stops.
The shutdown button turns red, indicating an error condition.
Note Do not select the red Shutdown button. The station is currently
in the red Shutdown state. In the red Shutdown state, the Centrifuge
module retains its current sample tube mapping. Selecting the red
Shutdown button causes the Centrifuge to go into a white Shutdown
state and the Centrifuge module cannot retain the current state of the
samples.
If a spin is in progress and the E-stop is not associated with Hettich unit,
the spin continues.
An event posts to the event log.
The LineMaster workspace status bar above the Centrifuge station
button turns red.
On the track light tower, the red light illuminates.
The Status Beacon is lit.
In software versions prior to 4.9, the Status Beacon is lit only when the
E-stop button is engaged.
Lifting the Robot from the Hettich Unit
If the robot is inside the Hettich unit, you may need to manually lift the
robot up from the Hettich unit before you can remove and clean sample
tubes, cassettes, buckets, and the Hettich unit.
To move the robot assembly, perform the following steps:
1. At the Centrifuge module, press the E-stop button.
BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to
Appendix A, Safety Information for recommended precautions when
working with biohazardous materials.