Centrifuge Module Interface Operating Instructions Troubleshooting
092D0001-02 Rev. A page 73
2. Open the robot enclosure access door and place the hand crank in the
hand crank receptacle.
Figure 7: Hand Crank Placement
3. Press and hold the Z-brake release and turn the hand crank to raise the
robot assembly.
4. After raising the robot assembly, remove the hand crank.
1. Hand crank
2. Hand crank receptacle
3. Z-brake release
CAUTION
Do not operate the Centrifuge module with the hand crank in the
module. Operating the Centrifuge module with the hand crank in place
can cause system damage. Store the hand crank in an easily accessible
location.