Centrifuge Module Interface Operating Instructions Maintenance
092D0001-02 Rev. A page 63
12. Using Hettich lubricating grease, lubricate the cylinder wall of the rotor
pins.
Figure 12: Rotor and Pins
Cleaning the Hettich Unit Chamber and Seals
To clean the Centrifuge chamber and seals, perform the following steps:
1. Inspect the outer and inner seals.
If a seal is damaged, call your local technical support provider for
replacement seals.
1. Rotor
2. Cylinder wall of the rotor pins
Cleaning Cycle As needed
Materials Needed Cleaning cloth
10% bleach solution
Hettich lubricating grease
Isopropyl alcohol