Maintenance Centrifuge Module Interface Operating Instructions
page 64 092D0001-02 Rev. A
2. Using an absorbent cloth and a 10% bleach and water solution, clean
around the inner and outer seals, the chamber, and the lid.
Ensure that you clean the area where the outer seal contacts the
underside of the lid.
3. Using a new absorbent cloth, dry the Centrifuge chamber, the seals,
and the lid.
4. Using Hettich lubricating grease, lightly grease the outer rubber seal of
the Centrifuge chamber.
Figure 13: Hettich Unit Seals
5. Replace the buckets.
6. Close the Hettich unit lid and turn the lid knob to the locked position.
1. Outer rubber seal
2. Inner seal