System Operation Centrifuge Module Interface Operating Instructions
page 38 092D0001-02 Rev. A
8. Press the robot computer power switch and wait approximately
2 minutes for the robot computer to power up.
Figure 5: Centrifuge Robot Computer Power Button
9. Close the panels.
10. On the LineMaster workspace station buttons, select Centrifuge.
The Details screen is displayed. The Centrifuge module is in a
Shutdown state.
11. On the States panel, select Online.
The interface transitions to the Online state.
The automation software and the Centrifuge module software
check communication links and safety features.
The robot surveys the Centrifuge module to check for sample tubes
in the Centrifuge.