Maintenance Centrifuge Module Interface Operating Instructions
page 54 092D0001-02 Rev. A
4. Remove the track access door to access the interface gate.
Figure 7: Interface Gate Tube Holder
5. Clean the interface gate tube holder:
a. Remove any debris and inspect for damage.
b. Dampen a soft cleaning cloth with 70% isopropyl alcohol and wipe
the tube holder.
You can use an alcohol prep pad instead of the damp cloth.
c. Wipe dry with a soft, clean, dry cleaning cloth.
d. Dampen a cleaning cloth with water and wipe the tube holder.
e. Wipe dry with a clean, dry cleaning cloth.
6. Replace the track access door.
7. Pull out the E-stop button.
8. To restore power to the interface gate, pull up the gate control button
on the associated divert gate.
1. Interface gate tube holder
2. Sample tube
3. Interface gate