Maintenance Centrifuge Module Interface Operating Instructions
page 58 092D0001-02 Rev. A
6. Using the T-handle 8-mm hex key included with your Centrifuge
module, remove the Hettich unit drawer locking screws.
Figure 9: Drawer Handle and Locking Screws
7. Verify that you have at least 1 x 1 m (3.3 x 3.3 ft) of clear space in front
of the drawer.
8. Use the drawer handle to pull the drawer, bringing it fully out of the
Centrifuge module until the drawer locks.
1. Drawer handle
2. Drawer locking screws
3. Drawer lock release
WARNING
Do not grasp the outside of Hettich centrifuge unit to pull out or push in
the drawer. Only use the Hettich drawer handle to pull the drawer out or
push the drawer back in place. Failure to use the drawer handle while
moving the Hettich centrifuge unit can pose a pinch or crushing hazard.