Centrifuge Module Interface Operating Instructions Troubleshooting
092D0001-02 Rev. A page 79
4. Turn the gripper release switch.
This opens the gripper and the sample tube releases.
Figure 9: Brake and Gripper Release
5. Turn the gripper release switch for approximately 10 seconds to lock
the gripper into the open position.
6. Inspect the gripper for debris or damage.
Clean the gripper if required, see Cleaning the Robot Gripper, page 47.
7. Close the robot access door.
8. Pull out the E-stop button.
9. Bring the Centrifuge module online.
1. Z-brake release
2. Gripper release switch