5-27
5
Instrument Set Up
4. Use the arrow keys to select
Authorized Operator.
5. Touch Select.
The Authorized operator screen
displays.
6. To permit access only by
authorized operators, touch
Enabled.
To allow all operators access to
the system, touch Disabled.
7. If you selected Enabled, see
Adding Operator IDs below to
add at least one operator.
If you selected Disabled, touch
Done 3 times to return to the
Select Ready screen.
CAUTION
If the instrument uses the operator
list sent by the LIS, do not
power down the system. If the
connector loses power, the
operator names are erased.
NOTE: The operator list sent by the
LIS overwrites an operator list
entered via the analyzer.
Adding Operator IDs
To add operator IDs, perform the
following steps:
1. At the Authorized operator
screen, touch Add operator.
2. Enter the new Operator ID.
Use the alpha keyboard to enter
text.
To enter numeric text, touch 123.
Instrument Settings
3. Touch Enter.
The Authorized Operator screen
displays indicating the Operator
ID and which functions the
operator can perform.
4. To edit this Operator ID, touch
Edit.