The MR image may show system-related artifacts/imaging errors
despite careful preparation.
If the same artifact/imaging error appears repeatedly, document
and submit it to Siemens Service.
CAUTION
RF-signal interference caused by incompatible accessories e.g.
patient monitoring devices!
Streaks and bright spots in the MR image
Use only accessories tested and approved for the MR system.
Keep the door to the examination room closed.
Vary the bandwidth of the MR sequence.
Whenever possible, use local coils for the MR examination.
CAUTION
Local variation in the sensitivity of local coils!
Continuous fluctuations in MR image brightness
Whenever possible, use a local coil with transmit
characteristics that are more suitable for the field of view
desired.
Use the normalization filter.
System-related artifacts/
imaging errors
Stripe artifacts
Variations in brightness
2 Safety
34
MR System | Operator Manual