CAUTION
Static and/or stationary brightness errors on the LCD monitor
due to aging!
Incorrect diagnosis
Scroll through the images to ensure that the MR image does
not show differences in brightness, spots, or cloudiness and
check bright objects for afterglow.
Keep your eyes always in central monitor position with a
vertical view angle towards the screen surface for best image
quality.
CAUTION
Inhomogeneous RF field!
Asymmetry of contrast in the MR image
Whenever possible, use a local coil with transmit
characteristics that are more suitable for the desired field of
view.
CAUTION
Spatial non-linearity of the gradient field and inhomogeneity of
the static magnetic field!
Pin-cushion and barrel-shaped distortions and/or loss of
signal in the margins of the MR image
Go through a distortion correction.
Position the region to be examined as close to the magnet
isocenter as possible.
Use phantoms for the control measurements.
Variations in signal and
contrast
Distortions/signal obliteration
along the edges
Safety 2
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
35
Print No. MR-01501G.621.02.02.02