EasyManua.ls Logo

Siemens MAMMOMAT Novation DR - Measuring the Line Resistance; Checking the Supply Voltages

Siemens MAMMOMAT Novation DR
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAMMOMAT
Novation
DR
SPB7-250.815.02 Page 4 of 6 Siemens AG
System Rev. 02 09.04 CS SD 24 Medical Solutions
3 - 4 Power up the MAMMOMAT Stand
Measuring the line resistance 3
1. Connect the line resistance meter to the mains input terminal on the stand.
2. Mains supply ON.
3. Carry out measurement.
4. Mains supply OFF.
To achieve the maximum output, the line resistance must not exceed the following values:
Checking the supply voltages 3
The correct supply voltages +5 V, +15 V and +28 V have already been tested in the fac-
tory. Only a visual check of the function is therefore needed.
1. Switch the system ON.
2. Check that the following LED indicators (top to bottom) on the mains input
converter (generator) are on:
Nominal Input Voltage
[VAC]
Line Resistance Ri
max
[c]
208 0,45
230 0,50
240 0,60
277 0,65
400 (2-phase) 0,85
Tab. 1 Measuring the line resistance
LED check
28 VP
5 VF
- 15 V
+ 15 V
28 VS
5 V
Tab. 2 LED check

Table of Contents

Other manuals for Siemens MAMMOMAT Novation DR

Related product manuals