SPB7-250.815.02 Page 8 of 12 Siemens AG
System Rev. 02 09.04 CS SD 24 Medical Solutions
5 - 8 Mechanical Checks
Checking the field light time 5
1. Switch on the field light by pressing the switch (1/Fig. 8) or by pressing the
compression foot switch (1/Fig. 9).
The field light shall go out automatically after a preset period of time (normally
20 s.).
Fig. 8 Field lightswitch
Fig. 9 Foot switch