Mechanical Checks 5 - 5
Siemens AG SPB7-250.815.02 Page 5 of 12 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
Mechanical Checks 5
Rotary motion 5
The rotary motion with the detector wing has to be checked.
1. System OFF.
The rotary motion and the vertical adjustment of the swivel arm system will be
blocked.
2. System ON, detector table installed and projection angle set to minimum with
presetting knob (1/Fig. 6).
3. Check the rotary motion of the swivel arm system by pressing buttons (2/Fig. 6)
one by one.
The rotation of the swivel arm system will stop when the button is released or
when the end position is reached (+135 degrees clockwise and -180 degrees
counterclockwise). The projection angle is shown on the display at the lower part
of the stand.