Detector installation 4 - 5
Siemens AG SPB7-250.815.02 Page 5 of 6 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
Mounting the grid 4
The materials needed for the grid installation can be found in the accessory material box.
1. Insert the left and right grid slider (1/Fig. 7) from the back into the rail.
2. Insert the left and right grid holder into the grid slider (2/Fig. 7).
3. Line up the grid to the front of the table, the front switch has to be engaged. Screw
the grid (3/Fig. 7) to the grid holder (4/Fig. 7). The marking has to be on top, text
must be readable.
4. Connect the cables (5/Fig. 7) for the magnification table, according to the labels,
on both sides.
Fig. 7 Grid installation
Make sure that the cables are connected according to the labels,
otherwise the D805 will burn up.