Checking dose settings 8 - 9
Siemens AG SPB7-250.815.02 Page 9 of 10 MAMMOMAT
Novation
DR
Medical Solutions Rev. 02 09.04 CS SD 24 System
4. Click Modify settings.
5. To adjust the
- H value,
go to Modify low dose scaling (The default values are 80, 70, 45) and enter
the new values, separated by a comma, e.g. 80,70,45 in the Low Dose Scal-
ing Value field and click Change Low Dose Scaling to save it.
- D value,
go to Modify high dose scaling (The default values are 100, 80, 60) and enter
the new values, separated by a comma, e.g. 100,80,60 in the High Dose Scal-
ing Value field and click Change High Dose Scaling to save it.
6. Proceed from where you have been referenced from.
Fig. 5 BRICK - Modify settings