Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
X-ray tube checks 45
Page 45 of 116
Medical Solutions
Oscilloscope diagram examples
Fig. 28: diagram_30kV_20mAs_Mo_Mo
Fig. 29: diagram_30kV_20mAs_W_Rh
30 kV 1 V/T (1 T <=> 5 kV) 30 kV 1 V/T (1 T <=> 5 kV)
130 mA 1 V/T (1 T <=> 40 mA) 166 mA 1 V/T (1 T <=> 40 mA)
160 ms 20 ms/T 128 ms 20 ms/T
Fig. 30: diagram_30kV_100mAs_Mo_Mo
Fig. 31: diagram_30kV_100mAs_W_Rh
30 kV 1 V/T (1 T <=> 5 kV) 30 kV 1 V/T (1 T <=> 5 kV)
112 mA 1 V/T (1 T <=> 40 mA) 132 mA 1 V/T (1 T <=> 40 mA)
0.9 sec 0.2 s/T 0.72 sec 0.2 s/T