EasyManua.ls Logo

Siemens MAMMOMAT Novation DR - Page 84

Siemens MAMMOMAT Novation DR
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84 Image Quality
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 84 of 116
Medical Solutions
Amplification Factor 2.52 Fix value for amplification, only
used when mode = OFF
Do not change
Amplification Mode ON See above Do not change
Set Direct Radiation
Black
ON Takes care that direct radiated area
is set to black
Do not change
Window Mode ON ON: Complete area of breast is
taken into account for the calcula-
tion of the window/center values
OFF: only compressed area is
taken into account for the calcula-
tion of the window/center values
Do not change
Edge Enhancement
Gain
0.3 Gain value for enhancement of the
local contrast
Do not change
Edge Enhancement
Kernel
15 Do not change
Edge Enhancement
LUT index
0 Currently not used; future plan to
have different edge enhancement
based on mean gray value in image
Do not change
LUT index 18 Separate LUTs to calculate the dis-
played values are delivered
Do not change
Harmonization Ampli-
fication Gain
0.7 Gain value for enhancement of the
global contrast
Do not change
Harmonization Kernel
Size
127 Gain value for enhancement of the
global contrast
Do not change
Factor make window
width wider
1.6 Defines range of gray values to be
displayed
Decrease if
image should
have more con-
trast; increase if
image should
have less con-
trast;
3 predefined
settings exist:
Normal, Low,
High
Parameter Default
Value
Description Changeable

Table of Contents