Siemens AG SPR8-220.101.04.03.02 MOBILETT Plus HP
07.05 CS PS 24
General 13
Page 13 of 64
Medical Solutions
1.9 Safety information and protective measures
WARNING
Risk of death, injuries or material damage.
Non-compliance can lead to death, injuries or material damage.
¹ When performing service work and tests, adhere to the
product-specific safety information in the documents.
¹ When performing service work and tests, adhere to the
general safety information contained in ARTD Part 2.
¹ Connect the MOBILETT Plus HP only to a mains power
supply outlet (receptacle) that corresponds to the instal-
lation requirements of VDE 0107 or country-specific reg-
ulations.
¹ Remove or install boards only when the generator is
switched OFF. Adhere to the ESD guidelines e .
¹ Checks and adjustments performed with radiation ON
are identified by the radiation warning symbol x . Dur-
ing these types of adjustments, radiation protective
clothing must be worn.
¹ Perform the protective conductor test according to
ARTD-002.731.17... upon completion of the work and
after all covers have been reattached.