Siemens AG SPR8-220.101.04.03.02 MOBILETT Plus HP
07.05 CS PS 24
General 13
Page 13 of 64
Medical Solutions
1.9 Safety information and protective measures
WARNING
Risk of death, injuries or material damage.
Non-compliance can lead to death, injuries or material damage.
¹ When performing service work and tests, adhere to the
product-specific safety information in the documents.
¹ When performing service work and tests, adhere to the
general safety information contained in ARTD Part 2.
¹ Connect the MOBILETT Plus HP only to a mains power
supply outlet (receptacle) that corresponds to the instal-
lation requirements of VDE 0107 or country-specific reg-
ulations.
¹ Remove or install boards only when the generator is
switched OFF. Adhere to the ESD guidelines e .
¹ Checks and adjustments performed with radiation ON
are identified by the radiation warning symbol x . Dur-
ing these types of adjustments, radiation protective
clothing must be worn.
¹ Perform the protective conductor test according to
ARTD-002.731.17... upon completion of the work and
after all covers have been reattached.

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals