28 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 28 of 64
Medical Solutions
2.2.2 Front wheels
Lock the rear wheels with the parking brake.
Lift up the front wheels using the tilt bar and jack them up with a piece of wood
(Fig.4/p.28).
SIM Secure mounting
SIM Ease of movement
Check that the wheels are securely mounted and that they turn freely. If necessary,
tighten the Allen screws (Fig.4/p.28).
PMP Cleaning
If the wheels need thorough cleaning, they must be removed:
Remove the Allen screws.
Remove the wheels (Fig.5/p.28).
Clean the wheels.
Remount the wheels.
Check the other set of wheels in the same way.
Fig. 4: Fig. 5: