26 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 26 of 64
Medical Solutions
2.2 Mechanical inspection
WARNING
Life-threatening electric shock hazard exists.
¹ Make sure the protective measures in chapter 1 have
been taken prior to opening the unit.
¹ Disconnect the system from the mains and disconnect
the batteries.
¹ (Protective measures for batteries and capacitors / p. 14)
.
2.2.1 Rear wheels
SIM Damage
Check the rear wheels for any damage.
Jack up one side of the unit with a piece of wood placed under the chassis
(Fig.1/p.26).
Secure the rear wheel on the other side with two blocks. Do not apply the parking
brake.
Remove the hub cap first and then the wheel (Fig.2/p.26).
SIM Drive belt
Check the condition and tension of the drive belt. The adjustment procedure is de-
scribed in the Service Instructions.
Fig. 1:
Fig. 2: