Siemens AG SPR8-220.101.04.03.02 MOBILETT Plus HP
07.05 CS PS 24
Inspection and Maintenance 33
Page 33 of 64
Medical Solutions
2.2.9 Mains cable
Fig. 11:
Pos. 1 Protective conductor
PMP Replace the mains cable every 2 years.
The mains cable must be replaced every two years.
SIM Damage
Pull the entire (6 m) mains cable out from the cable winder.
Check the mains cable and plug for damage. If necessary, replace them. See section
"Replacement of mains cable / cable winder" in the Service Instructions.
PMP Cleaning
Clean the mains cable using a paper towel dampened with a lukewarm solution of wa-
ter and a household cleaning agent.
Pull the brake handle for the cable upwards.
SIM Cable winder
Move the cable up and down while rewinding it, in order to avoid any slack, until the ca-
ble is completely rolled up. Check the proper functioning of the cable winder and brake.
SIM Protective conductor
Disconnect the power supply plug.
Check the protective conductor connection on the plug. The protective conductor must
be securely attached and must be slightly longer than the other two wires. This applies
for all plug types.
Close the plug after inspecting it.

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals