2.11 DAP measuring system (optional)
• Press the test button on the display. The value on the display should be in the range
80-120 µGym
2
(DAP resolution 0.1 µGym
2
) or 8-12 µGym
2
(when using a chamber of
high sensitivity, i.e., DAP resolution 0.01 µGym
2
). If not, calibrate the system according
to the Service Instructions.
X
• Check the function of the DAP measuring system by releasing an exposure. Check if a
measured value is shown on the display.
SIE Function of DAP measuring system