Siemens AG SPR8-220.101.04.03.02 MOBILETT Plus HP
07.05 CS PS 24
Inspection and Maintenance 59
Page 59 of 64
Medical Solutions
2.17 Protective ground/earth measurement
Perform the protective conductor test, with all system covers closed, according to
ARTD-002.731.17... (safety-related regulations on installation and maintenance). The
measured values are to be entered in the maintenance protocol. The protective con-
ductor resistance must not exceed 0.2 Ohms.