58 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 58 of 64
Medical Solutions
2.16 Checking the manual exposure interrupt
The user must be able to interrupt the exposure at any time.
Set 70 kV, 160 mAs.
X
Release an exposure and immediately interrupt it by letting go of the exposure switch.
The message ”USE 20” appears on the display indicating that the exposure release
switch was interrupted during the exposure. In addition, 10 short "beep" signals sound.
The message can be erased by pressing the ”kV+” button.
SIE Message "USE20"
SIE 10 short beep signals
NOTE
All covers must be remounted.