62 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 62 of 64
Medical Solutions
2.20 Checking the kV and mAs displays
• Check whether the green "ready" lamp lights up approximately 2 minutes after switch-
ing the system on. If it doesn’t light up, an error message will appear on the display. Re-
fer to the chapter "Display messages" in the Service Instructions.
• Check whether the kV and mAs values are clearly visible from the operator’s position.
Determine whether all exposure parameters are correctly displayed.
PMF "Ready" indicator
SIE Visibility of displays
SIE Correct displays