Siemens AG SPR8-220.101.04.03.02 MOBILETT Plus HP
07.05 CS PS 24
Inspection and Maintenance 27
Page 27 of 64
Medical Solutions
NOTE
Under normal operating conditions, the drive belt does not require
any adjustment throughout the lifetime of the system.
Fig. 3:
SIM Brake lining
Check the brake lining (Fig.3/p.27). If the lining is worn down to less than 1 mm, it
must be replaced. Refer to the section on ”Adjusting the hand brake/parking brake” in
the Service Instructions.
Remount the wheel.
SIM Secure mounting
SIM Ease of movement
Check that the wheel is securely mounted and that it turns freely.
Check the other wheel following the same procedure.