44 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 44 of 64
Medical Solutions
* The measurement inaccuracy of the respective measurement device must be sub-
tracted from these limit values.
Setting
• Proceed according to section "P10 - Fast adaptation" in the Service Instructions in or-
der to adjust excessively high deviations.
PMF kV accuracy 52 kV, 50 mAs (B)
PMF kV accuracy 81 kV, 20 mAs (B)
PMF kV accuracy 133 kV, 12.5 mAs (B)
• Connect the power supply plug and switch the main switch to position "M"
X
• Set the following exposure parameters and release an exposure:
• Disconnect the power supply plug, roll up the cable and switch the main switch to posi-
tion "B".
PMF kV accuracy 81 kV, 20 mAs (M)
kV Meter
selection
Exposure parameter Limit values* to be maintained
DC voltage
52 kV 50 mAs 49.0 - 55.0 kV
81 kV 20 mAs 77.0 - 85.0 kV
133 kV 12.5 mAs 126.4 - 139.6 kV
kV Meter
selection
Exposure parameter Limit values* to be maintained
DC voltage 81 kV 20 mAs 77.0 - 85.0 kV