44 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 44 of 64
Medical Solutions
* The measurement inaccuracy of the respective measurement device must be sub-
tracted from these limit values.
Setting
Proceed according to section "P10 - Fast adaptation" in the Service Instructions in or-
der to adjust excessively high deviations.
PMF kV accuracy 52 kV, 50 mAs (B)
PMF kV accuracy 81 kV, 20 mAs (B)
PMF kV accuracy 133 kV, 12.5 mAs (B)
Connect the power supply plug and switch the main switch to position "M"
X
Set the following exposure parameters and release an exposure:
Disconnect the power supply plug, roll up the cable and switch the main switch to posi-
tion "B".
PMF kV accuracy 81 kV, 20 mAs (M)
kV Meter
selection
Exposure parameter Limit values* to be maintained
DC voltage
52 kV 50 mAs 49.0 - 55.0 kV
81 kV 20 mAs 77.0 - 85.0 kV
133 kV 12.5 mAs 126.4 - 139.6 kV
kV Meter
selection
Exposure parameter Limit values* to be maintained
DC voltage 81 kV 20 mAs 77.0 - 85.0 kV

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals