46 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 46 of 64
Medical Solutions
NOTE
The signal contains a high frequency of 20-40 kHz, which can lead
to measurement problems with the oscilloscope (aliasing effect).
Always use an oscilloscope with a high sampling rate.
X Calculate the product of the tube current x the exposure time for the following exposure
parameters:
* The measurement inaccuracy for the measurement device must be subtracted from
these limit values.
Determine whether the mAs product (tube current + reactive current) x exposure time
lies within the indicated tolerances.
Setting
Proceed according to section "P10 - Fast adaptation" in the Service Instructions to elim-
inate slight deviations.
PMF mAs accuracy 40 kV, 5 mAs (B)
PMF mAs accuracy 81 kV, 2 mAs (B)
PMF mAs accuracy 133 kV, 10 mAs (B)
Connect the power supply plug and switch the main switch to position "M".
X
Set the following exposure parameters and release an exposure:
Disconnect the power supply plug, roll up the cable and switch the main switch to posi-
tion "B".
PMF mAs accuracy 81 kV, 2 mAs (M)
Set on the control and display panel Acceptable mAs value*
kV mAs
40 5 4.7 - 5.3
81 2 1.9 - 2.1
133 10 9.5 - 10.5
Set on the Control and display panel Acceptable mAs value*
kV mAs
81 2 1.9 - 2.1

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals