48 Inspection and Maintenance
MOBILETT Plus HP SPR8-220.101.04.03.02 Siemens AG
07.05 CS PS 24
Page 48 of 64
Medical Solutions
If the value for the fluctuation coefficient calculated is higher than 0.045, the reason for
the deviation must be determined and corrective measures taken. Refer to section "P10
- Fast adaptation" in the Service Instructions.
QSQ Fluctuation coefficient C
Steps for calculating C
Example:
The values used in the example are not real values.
Release 10 exposures and record the values measured.
Add the measured values:
Σ = 1.01 + 1.02 + 1.03 + 1.04 + 1.03 + 1.02 + 1.02 + 1.01 + 1.03 + 1.04 => Σ = 10.25
Calculate the average value:
Fig. 24:
Calculate the standard deviation:
Fig. 25:
Exposure
(n=10)
Measured values
(xi)
11.01
21.02
31.03
41.04
51.03
61.02
71.02
81.01
91.03
10 1.04