Technical Data
Operator Manual
SPR8-230.621.01.03.02
88 of 106
Guidance and manu-
facturer's declaration -
electromagnetic
emission
The Mobilett XP is intended for use in the electromagnetic environment specified
below. The customer or the user of the Mobilett XP should assure that it is used
in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
The Mobilett XP uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby elec-
tronic Equipment.
RF emissions
CISPR 11 Class B
The Mobilett XP is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emmisions
IEC 61 000-3-3 Not applicable
Voltage fluctuations/
flicker emissions
IEC 61 000-3-3
Not applicable
The Mobilett XP has a rated input current of more than
16 A per phase.
Warning:
The equipment or system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the equipment or
system should be observed to verify normal operation in the configuration in
which it will be used.

Table of Contents

Other manuals for Siemens MOBILETT XP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT XP and is the answer not in the manual?

Siemens MOBILETT XP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT XP
CategoryMedical Equipment
LanguageEnglish

Related product manuals