Technical Data
MOBILETT
SPR8-230.621.01.03.02 89
of 106
Guidance and manu-
facturer's declaration -
electromagnetic
immunity
The Mobilett XP is intended for use in the electromagnetic environment specified
below. The customer or the user of the Mobilett XP should assure that it is used
in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61 000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61 000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Surge
IEC 61 000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Voltage dips, short
interruptions and volt-
age variations on power
supply input lines
IEC 61 000-4-11
<5% U
T
(>95%
dip in U
T
) for 0.5
cycle
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
<5% U
T
(>95%
dip in U
T
) for 5
cycles
n.a. Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the Mobilett XP
requires continued operation during power
mains inter-ruptions, it is recommended
that the Mobilett XP be powered from an
uninterruptible power supply or a battery.
The Mobilett XP has a rated input current of
more than 16 A per phase.
Power frequency
(50/60 Hz)
magnetic field
IEC 61 000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.

Table of Contents

Other manuals for Siemens MOBILETT XP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT XP and is the answer not in the manual?

Siemens MOBILETT XP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT XP
CategoryMedical Equipment
LanguageEnglish

Related product manuals