EasyManua.ls Logo

Siemens POLYMOBIL III - Page 33

Siemens POLYMOBIL III
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Notes on trouble-shooting 4 - 21
Siemens AG RXB8-115.061.02 Page 21 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Fig. 4
Evaluation:
Measure the deviations between light field edges and radiation field edges on all four
sides (X1,X2,Y1, Y2) as shown in Fig. 4.
Determine the total deviations in the X and Y direction (ignore the +/- signs).
Both the length deviation ( Σ Y) and the width deviation ( Σ X) must be less than 1.6 cm.
X
1
Y
1
Y
2
X
2
Film
Light field
Radiation field

Other manuals for Siemens POLYMOBIL III

Related product manuals