POLYMOBIL III RXB8-115.061.02 Page 22 of 24 Siemens AG
Rev. 01 10.95 CS PS 24 Medical Solutions
4 - 22 Notes on trouble-shooting
• If the deviation is higher, loosen the 4 Allen screws slightly (arrows/Fig. 5/Fig. 6) and
move the collimator accordingly.
Then, tighten the screws at the collimator (arrows/Fig. 5/Fig.6) again.
• Repeat the check and, if necessary, adjust the collimator again until the deviation
between the light field and the radiation field is within the admissible tolerance
(< 1,6 cm).
Fig. 5 Fig. 6