SITRANS FM100
A5E48762204-AA page 41
11. Status
The electromagnetic flowmeter can detect and display various device or
application errors.
If there is a status or error message, the STATUS symbol in the display
alternately flashes orange / red. To call up the status / error information, the
status key must be pressed, then the status window that appears then lists all the
messages that have accumulated up to this point in time. By pressing the ⋅⋅
key, the user confirms the knowledge of the displayed errors, the status memory
is cleared and the status window is closed. If one of the displayed errors persists,
it will be reported again by flashing the status icon.
The following status / error messages are generated:
Display text
Description
Debugging
Empty Pipe
Measuring tube is not completely
filled with medium or medium
with too low conductivity is used.
Check the filling of the
measuring circuit or
conductivity of the medium
(> 20 μs / cm)
Temp Sens Error
Error in the temperature
measuring circuit
Repair by Siemens Service
necessary
Meas saturated
Flow measuring circuit
overdriven
Reduce flow rate
No Subslave Internal hardware error
Repair by Siemens Service
necessary
Simulation
Simulation function active
-