SITRANS FM100
page 54 A5E48762204-AA
17. Appendix
17.1 IO-Link process data structure
Process data length: 9 bytes
Byte
no.
Data
Bit count
Format
Factor
Range
Value
0 - 2
Flow
24 Bit
IntegerT
1/1000
+/-8388,608
L/min
3 - 6
Volume
32 Bit
FloatT
+/-1,4*10
-45
... +/-3,4*10
+38
L
7 - 8
Temperature
12 Bit
IntegerT
1/10
+/-204,8
°C
reserved
1 Bit
BooleanT
reserved
1 Bit
BooleanT
Status OUT1
1 Bit
BooleanT
Status OUT2
1 Bit
BooleanT
0
1
2
3
4
5
6
7
0
1
2
3
4
5
6
7
Flow (24 Bit, IntegerT)
Byte 0
Byte 1
0
1
2
3
4
5
6
7
Byte 2
0
1
2
3
4
5
6
7
0
1
2
3
4
5
6
7
Volume (32 Bit, FloatT)
Byte 3
Byte 4
0
1
2
3
4
5
6
7
Byte 5
0
1
2
3
4
5
6
7
Byte 6
0
1
2
3
4
5
6
7
0
1
2
3
4
5
6
7
Temperature (12 Bit, IntegerT)
Byte 7
Byte 8