9.3.10 Frequency (3.10)
Se
ts the country speci󹪝c settings for radar signals.
Setting
Mode1
EU, Albania, Andorra, Azerbaijan, Australia, Belarus, Bosnia
and
Herzegovina, Canada, Liechtenstein, Moldavia, Monaco,
Montenegro, New Zealand, Northern Macedonia, Norway,
San Marino, Saudi Arabia, Serbia, Switzerland, Turkey, Uk‐
raine, United Kingdom, USA
Mode2
Kazakhstan
Mode3
Brazil, South Korea, Taiwan, Thailand, Japan
Mode4
India, Malaysia
Default Mode1
9.3.11 Audit trail (3.11)
9.3.11.1 Con󹪝gur
ation change counter (3.11.1)
Displays number of times device con󹪝guration or calibration changed, locally or via engineering
system.
9.3.12 Firmware update (3.12)
9.3.12.1 Data transfer (3.12.1)
Analog output set to 22.8mA during 󹪝rmware update.
9.3.12.2 Firmware update (3.12.2)
9.4 Communication (4)
The following parameters relate to device Communication.
9.4.1 Polling address (4.1)
Sets poll ID on a HART network.
For point-to-point con󹪝gurations, the standard address is zero (0). For multidrop
con󹪝gurations, use a non-zero HART address.
Parameter assignment
9.4Communication (4)
SITRANS LR500 series with mA/HART
Operating Instructions, 03/2024, A5E51099898-AA 145