Loading...
Variable Value
Max. scale interval n
LC
3000/4000
Minimum scale interval V
min
For E
max
= 2.3; 4.5; 9.1 t
For E
max
= 13.6; 18; 23; 27; 34; 45; 68; 91;113 t
E
max
/ 10000; 15000
E
max
/ 10000
Combined error F
comb
≤± 0.0230 % Cn for C3
≤± 0.0175 % Cn for C4
Min. initial loading E
min
0 kg
Maximum lateral load L
lq
100% E
max
Maximum working load L
u
150% E
max
Break load L
d
300% E
max
Rated measuring path h
n
For E
max
= 2.3 t:
For E
max
= 4.5 t:
For E
max
= 9.1 t:
For E
max
= 13.6 t:
For E
max
= 18 t:
For E
max
= 23 t:
For E
max
= 27 t:
For E
max
= 34 t:
For E
max
= 45 t:
For E
max
= 68 t:
For E
max
= 91 t:
For E
max
= 113 t:
0.5 mm
0.6 mm
1.1 mm
0.5 mm
0.5 mm
0.5 mm
0.6 mm
0.5 mm
0.5 mm
0.5 mm
0.9 mm
0.9 mm
Supply voltage U
SR
Maximum supply voltage 18 V DC
5 ... 12 V DC
Rated characteristic value C
Rated
3.0 ± 0.008 mV/V
Tolerance of zero signal D
o
≤ ± 1.0% C
n
Input resistance R
i
700 ± 7 Ω
Output resistance R
o
703 ± 4 Ω
Insulation resistance R
is
5000 MΩ at 50 V DC
Rated temperature range B
tn
-10 to +40 °C
Operating temperature range B
tu
-35 to +65 °C
Storage temperature range B
ts
-40 to +80 °C
Sensor material Steel, nickel-plated
Degree of protection according to EN 60 529 IP67
Cable connection cable length Four-core, shielded PU cable
9 m
Technical data
9.1 Functional data
SIWAREX WL200
Operating Instructions, 12/2021, A5E02199611B-13 69

Table of Contents