9.2 Energy limitation parameters
Load cell type Type of protection
ia and ic tc
1)
nA
1)
WL230 BB-S SA
WL230 SB-S SA
WL230 SB-S CA
WL250 ST-S SA
WL260 SP-S SA
WL260 SP-S SB
WL260 SP-S SC
WL270 CP-S SA
WL270 CP-S SB
WL290 DB-S CA
U
i
= 20 V DC
I
i
= 420 mA
P
i
= 4.2 W
L
i
2)
= 40 µH
C
i
2)
= 12 nF
-20 °C ≤ T
a
≤ +40 °C
Load cell connected by a sin‐
gle associated apparatus:
U
i
= DC 16 V
I
i
= 420 mA
P
i
= 4.2 W
U
n
= 20 V DC
-20 °C ≤ T
a
≤ +40 °C
U
n
= 20 V DC
-20 °C ≤ T
a
≤ +40 °C
WL270 CP-S SC
WL270 K-S CA
- U
n
= 20 V DC
-20 °C ≤ T
a
≤ +40 °C
U
n
= 20 V DC
-20 °C ≤ T
a
≤ +40 °C
WL280 RN-S SA
1)
P
i
= 1.3 W
-30 °C ≤ T
a
≤ +40 °C
P
i
= 1.25 W
-30 °C ≤ T
a
≤ +50 °C
P
i
= 1.2 W
-30 °C ≤ T
a
≤ +60 °C
P
rated
= 0.25 W
U
n
= 15 V DC
-30 °C ≤ T
a
≤ +60 °C
U
n
= 15 V DC
-30 °C ≤ T
a
≤ +60 °C
U
i
= 19.1 V DC
i
i
= 200 mA
L
i
= 2.7 µH + 1.3 µH/m
Load cell cable
C
i
= 0.0001 nF + 0.057 nF/m
Load cell cable
1)
Take measures to ensure that the supply voltage (including transients) does not rise above 140% of the
rated voltage.
2)
With a load cell cable length of 200 feet / 60.96 m
See also
Basic safety instructions (Page 25)
Basic safety instructions (Page 21)
9.3 Approval to OIML R60
Most SIWAREX load cells are approved for use in scales requiring ocial calibration of Class III,
EN 45501.
This is determined by the specications of the respective type.
Technical data
9.3 Approval to OIML R60
SIWAREX WL200
70 Operating Instructions, 12/2021, A5E02199611B-13