WARNING
Insucient ventilation
The device may overheat or start burning in the case of insucient ventilation.
• Ensure sucient ventilation between the devices when installing in control cabinets. The
heat sinks at the rear must remain free for air circulation.
• Make sure during operation that the permissible ambient temperature range is always
observed (seeTechnical specications (Page51)).
CAUTION
Strong vibrations
Strong vibrations could loosen connections or damage sensors, resulting in free passage of the
sample gas into the environment.
Even weaker vibrations inuence the result!
The analyzer must therefore only be used at a location which is free of vibration.
Please observe the data in the Technical specications.
3.3 Analyzers in hazardous areas
DANGER
Explosion hazard
If a ammable or ignitable atmosphere exists, plugs must never be disconnected or lamps/
fuses replaced when the analyzer is supplied with power.
Assembly/installation
3.3Analyzers in hazardous areas
Series 6 and ULTRAMAT 23
26 Compact Operating Instructions, 01/2024, A5E45779144002-AB