19" rack unit with
hoses
incl. with option‐
al order option
E20
19" rack unit
with hoses
ONLY for order
option E21 *
19" rack unit
with pipes
Bench-top unit
with hoses
(not for Ex appli‐
cations)
Analyzer chamber O-
ring
FKM x x x
Condensation trap at
gas inlet
PA6 (polyamide) PE
(polyethylene)
x
* Wall strengths of the components used for E21 all ≥ 1mm
7.2 General technical data series 6
The general technical specications apply to all analyzer versions unless they are not dened
otherwise in the technical specications of the analyzer versions.
The technical specications listed in this document are an excerpt that includes all safety-
related data. The complete technical specications can be found in the respective operating
instructions of your analyzer.
Enclosure
Position of use of the analyzer Front panel vertical
Degree of protection IP20 according to EN / IEC60529
Electrical characteristics
EMC interference immunity
(with safety extra-low voltage (SELV) with
safe isolation)
In accordance with standard requirements of NAMUR NE21or EN / IEC 61326
Electrical safety In accordance with EN / IEC 61010-1,
overvoltage category III
Auxiliary power supply 100V AC to 120V AC, 50/60Hz or
200V AC to 240V AC, 50/60Hz
Climatic conditions
Permissible ambient temperature
• During operation
– CALOMAT 6E, article no. 7MB252x-xGAxx
– All other analyzers
• During transportation and storage
• Pollution degree
+5to+55°C
+5to+45°C
-20°Cto+ 60°C (-4 °Fto+140 °F)
2 according to IEC 60664-1
Vibrations
• During operation
• During transportation and storage
None
Maximum vibration 5 m/s
2
Permissible ambient pressure 700to1200 hPa absolute
Technical specications
7.2General technical data series 6
Series 6 and ULTRAMAT 23
54 Compact Operating Instructions, 01/2024, A5E45779144002-AB