Technical speci󹪝cations
7
7.1 Technical speci󹪝cations ULTRAMAT 23
General information
Measured components Maximum of 4
Measuring ranges 2 per component, freely adjustable between smallest and largest measur‐
ing range
Characteristics Linearized
Operator panel LCD with LED backlighting and contrast control, 80 characters (4lines à
20characters); function keys
Operating position Front panel vertical
Vibrations
During operation
During transport
No vibrations
Maximum vibration 5 m/s
2
Warm-up phase (function check activated) 30 minutes if no UV module is installed
60 minutes if UV module is installed
Enclosure
Weight Approx. 10 kg (22 lbs.) The weight varies according to the ordered variant.
Degree of protection IP40 in accordance with EN / IEC60529 for the versions 7MB235x (TÜV)
IP20 in accordance with EN / IEC 60529 for the versions 7MB233x (standard)
Electrical characteristics
Electrical safety In accordance with EN / IEC 61010-1,
overvoltage category II
EMC interference immunity
(with safety extra-low voltage (SELV) with safe
isolation)
In accordance with standard requirements of NAMURNE21or
ENIEC 61326-1
Auxiliary power supply 100 V AC, +10%/-15%, 50 Hz,
120 V AC, +10%/-15%, 50 Hz,
200 V AC, +10%/-15%, 50 Hz,
230 V AC, +10%/-15%, 50 Hz,
100 V AC, +10%/-15%, 60 Hz,
120 V AC, +10%/-15%, 60 Hz,
230 V AC, +10%/-15%, 60 Hz
Power consumption Max. 60VA depending on ordered version
Fuse values 200V/230VAC; T 630mA/L 250V
see inscription on rear of device
100V/120VAC; T 1.25A/L 250V
see inscription on rear of device
Electrical inputs and outputs
Series 6 and ULTRAMAT 23
Compact Operating Instructions, 01/2024, A5E45779144002-AB 51