Parts in gas path wetted by sample gas
Sample chamber
Body
Lining
Gaskets (O-rings)
Window
Alu
Alu or Ta
FKM or FFKM
CaF
2
, adhesive E353
Special applications
Pipe gas path Gas feedthrough Ti or stainless steel1.4571 or Hastelloy C22
Pipe Ti or stainless steel1.4571 or Hastelloy C22
Sample chamber
Body
Lining
Gaskets (O-rings)
Window
Ti or stainless steel1.4571 or Hastelloy C22
Ta or Alu
Material according to speci󹪝cation
CaF
2
; without adhesive
Further designs upon request
Options
Flowmeter Measurement pipe Duran glass
Variable area Duran glass
Suspension boundary PTFE
Angle pieces FKM
Pressure switch sample gas Diaphragm FKM
Enclosure PA 6.3 T
7.4 Technical speci󹪝cations OXYMAT 6E, OXYMAT channel in
OXYMAT/ULTRAMAT 6E and OXYMAT 61
Enclosure
Weight Approx. 13kg
Dimensions 483 mm x 385mm x 172 mm (W x D x H)
Electrical characteristics
Power consumption (analyzer) Approx. 45VA
Fuse ratings
100 ... 120V
200 ... 240V
1T/250
0.63T/250
Gas inlet conditions
Permissible sample gas pressure 800 to 1200hPa absolute (with external reference gas supply),
atmospheric pressure (with built-in pump)
Sample gas 󹪞ow 18 to 60 l/h (0.3 to 1 l/min)
Technical speci󹪝cations
7.4Technical speci󹪝cations OXYMAT 6E, OXYMAT channel in OXYMAT/ULTRAMAT 6E and OXYMAT 61
Series 6 and ULTRAMAT 23
56 Compact Operating Instructions, 01/2024, A5E45779144002-AB