EasyManua.ls Logo

Siemens ULTRAMAT 6 Series - Checking for Leaks

Siemens ULTRAMAT 6 Series
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.3 Checking for leaks
Checking the sample gas path for leaks
Carry out a leak test prior to commissioning or following each maintenance measure which
a󹪜ects the sensor or the gas path. This is carried out as follows:
1. Connect the sample gas outlet to a relative pressure manometer (e.g. U-tube manometer
with measuring range 0 to 150hPa, resolution 0.1hPa).
2. Apply air to the sample gas inlet.
As soon as an overpressure of approx. 100 hPa has been reached (measure using manometer
at sample gas outlet), block the sample gas inlet.
3. Wait approx. one minute until the enclosed air has become adapted to its environmental
conditions. Then record the pressure.
4. Wait for a while and read the pressure once again. The gas path is su󹪟ciently tight when the
pressure drop does not exceed a maximum value.
Di󹪜erent criteria are in e󹪜ect for the parameters (wait time and pressure drop) listed in item
4 due to the di󹪜erent designs of the analyzers. You can 󹪝nd the relevant values for each
analyzer in the operating instructions of the respective unit.
Note
It must be guaranteed during this measurement that the temperature of the gas path and sensor
remains constant.
Commissioning
5.3Checking for leaks
Series 6 and ULTRAMAT 23
46 Compact Operating Instructions, 01/2024, A5E45779144002-AB

Table of Contents

Related product manuals