7.3 Technical specications ULTRAMAT 6E and ULTRAMAT channel in
ULTRAMAT/OXYMAT 6E
Enclosure
Weight Approx. 15kg with one IR channel
approx. 21 kg with two IR channels
Dimensions 483 mm x 385mm x 177 mm (W x D x H)
Electrical characteristics
Power consumption
• 1-channel unit
• 2-channel unit
Approx. 40VA
approx. 70VA
Fuse values
• 1-channel unit (7MB2121, 7MB2127, 7MB2124, 7MB2126)
– 100 to 120V
– 200 to 240V
• 2-channel unit (7MB2123, 7MB2128, 7MB2124, 7MB2126)
– 100 to 120V
– 200 to 240V
1T/250
0.63T/250
1.6T/250
1T/250
Gas inlet conditions
Permissible sample gas pressure 600to1500 hPa absolute,
700to1300hPa absolute with installed pressure switch
Sample gas ow 18to90 l/h (0.3 to 1.5l/min)
Sample gas temperature 0to50°C
Sample gas humidity <90%RH (relative humidity), non-condensing or
dependent on measuring task
Parts in gas path wetted by sample gas
Standard devices
Gas path with tubes Gas feedthrough Stainless steel1.4571
Hose FKM
Hose coupling PA 6
Sample chamber
• Body
• Lining
• Gaskets (O-rings)
• Window
Alu
Alu
FKM or FFKM
CaF
2
, adhesive E353
Pipe gas path Gas feedthrough Ti or stainless steel1.4571
Pipe Ti or stainless steel1.4571
Technical specications
7.3Technical specications ULTRAMAT 6E and ULTRAMAT channel in ULTRAMAT/OXYMAT 6E
Series 6 and ULTRAMAT 23
Compact Operating Instructions, 01/2024, A5E45779144002-AB 55