Commissioning
5
5.1 Safety instructions
WARNING
Hazardous contact voltage
Risk of injury through hazardous contact voltage when the device is open or not completely
closed.
The degree of protection specied on the nameplate or in Technical specications (Page51)
is no longer guaranteed if the device is open or not properly closed.
• Make sure that the device is securely closed.
WARNING
Use of ammable and toxic gases
If ammable gases including occasionally explosive gases (Ex Zone 1) are discharged from
ULTRAMAT 23 equipment, a solenoid valve is part of the containment system and the zero-
point calibration (AUTOCAL) is not performed with air instead of inert gas, a daily AUTOCAL has
to be performed. In this way, a possible leakage at the solenoid valve gasket can be detected
through zero point deviation during calibration.
If a zero point deviation occurs for this conguration, the device must be taken out of operation
immediately and a leak test must be performed, in particular of the solenoid valve.
In hazardous areas (Zone 2), measuring of ammable gases is only permissible with the option
-E21.
CAUTION
Loss of degree of protection
Damage to device if the enclosure is open or not properly closed. The degree of protection
specied on the nameplate or in Chapter "Technical specications (Page51)" is no longer
guaranteed.
• Make sure that the device is securely closed.
Series 6 and ULTRAMAT 23
Compact Operating Instructions, 01/2024, A5E45779144002-AB 43