Please also observe the following for correct operation:
Connect the analyzer's enclosure to the equipotential bonding.
Connect the signal lines to the Sub-D plugs at the rear of the device.
Refer to the ELAN interface description (Order No. C79000-B5200-C176 German, C79000-
B5276-C176 English) for details on the interface cable.
4.2.3 PROFIBUS DP
1
2
3
4
5
6
7
8
9
68%');FRQQHFWRU
352),%86'3
'*1'
&1751
&1753GLUHFWLRQFRQWURO
939
1&QRWDVVLJQHG
1&QRWDVVLJQHG
5['7['1$
5['7['3%
Figure4-2 PROFIBUS DP connection
CAUTION
Pin 7 is connected to the data line. No consumers must be connected.
The cable shield must be connected over a larger surface in the plug housing.
Failure to comply with this can result in personal injury or damage to property.
Connecting
4.2Electrical connection
Series 6 and ULTRAMAT 23
34 Compact Operating Instructions, 01/2024, A5E45779144002-AB