Time response
Warm-up time <30min, maximum accuracy is achieved after about 2 hours
Delayed display T
90
(incl. dead time) Approx. 35s
Damping (electronic time constant) 0 to 100 s, adjustable
Dead time (at 1 l/min) Approx. 34 s (because of di󹪜usion to sensor)
7.8 Technical speci󹪝cations FIDAMAT 6
Enclosure
Weight Approx. 23kg
Dimensions (W x D x H) 483mm x 457mm x 160 mm
Electrical characteristics
Power consumption Approx. 150VA during operation
Approx. 350VA in the warm-up phase
Fuse values
100 to 120V
200 to 240V
4.0T/250
2.5T/250
Gas inlet conditions
Sample gas pressure
With integrated pump
Without pump
600 ... 1100hPa
<2000hPa
Sample gas temperature 0...200°C
Sample gas humidity < 90% relative humidity
Pressures and 󹪞ow (FIDAMAT 6 with pump)
Gases Inlet pressure hPa
(rel.)
Operating pressure
hPa (rel.)
Flow across FID
ml/min
Flow through by‐
pass
ml/min
Without pump
start-up
With pump start-up
Combustion
gas
2000 ... 4000 1000 ± 20 ~ 12 -
Combustion
air
2000 ... 4000 420 ± 20 500 ~ 320 ~ 500
Sample gas atmospheric - 500 ± 2 ~ 3 ~ 1000
Zero gas 2500 ... 3000 - 1500 ± 2 ~ 3 ~ 1000
Calibration gas 2500 ... 3000 - 1500 ± 2 ~ 3 ~ 1000
Technical speci󹪝cations
7.8Technical speci󹪝cations FIDAMAT 6
Series 6 and ULTRAMAT 23
60 Compact Operating Instructions, 01/2024, A5E45779144002-AB